Title: An Improved Assay Method For The Estimation Of Ticagrelor Hydrochloride By Reverse Phase Liquid Chromatography
Authors: Ambasana, M.A.
Kapuriya, N.P.
Mangtani, K.M.
Ladva, K.D.
Keywords: Ticagrelor HCl; Antiplatelet agent; HPLC; Validation
Ticagrelor HCl; Antiplatelet agent; HPLC; Validation
Issue Date: 1-May-2016
Publisher: International Journal Of Pharmaceutical Sciences And Research
Citation: Ambasana, M. A., Kapuriya, N. P., Mangtani, K. M., & Ladva, K. D. (2016). An improved assay method for the estimation of ticagrelor hydrochloride by reverse phase liquid chromatography,International Journal Of Pharmaceutical Sciences And Research7(5)2009-2014.
Abstract: The current investigation was carried out to develop and validate a fairly simple, accurate, precise, reproducible and robust RP-HPLC method for the estimation of Ticagrelor Hydrochloride. The separation was achieved using Agilent Infinity 1220, Infinity Fast-LC (Pressure limit up to 600 bars) with auto sampler and PDA detector. The Chromatographic analysis was performed on ZORBAX Eclipse Plus 300SB C18 (250 x 4.6mm, 5.0 micron, PN 880995-902) column. Mobile phase consist of (A) Acetonitrile: (B) 20mM Potassium dihydrogen ortho phosphate buffer (40:60 v/v) at a flow rate of 1.0 ml/min. The method showed linear in the mentioned concentrations having line equation y = 22.848x + 1.3214 with correlation coefficient R2 of 0.9995. The recovery values for Ticagrelor ranged from 99.63% to 100.34%. The % RSD was 0.49% and 0.54%, respectively for intraday and interday precision. The limit of detection and limit of quantification were 0.05μg/mL and 0.20μg/mL respectively. Newly developed method was statistically validated for accuracy, precision, linearity and solution stability; hence it is directly applicable for the estimation of Ticagrelor up to trace level in routine analysis.
The current investigation was carried out to develop and validate a fairly simple, accurate, precise, reproducible and robust RP-HPLC method for the estimation of Ticagrelor Hydrochloride. The separation was achieved using Agilent Infinity 1220, Infinity Fast-LC (Pressure limit up to 600 bars) with auto sampler and PDA detector. The Chromatographic analysis was performed on ZORBAX Eclipse Plus 300SB C18 (250 x 4.6mm, 5.0 micron, PN 880995-902) column. Mobile phase consist of (A) Acetonitrile: (B) 20mM Potassium dihydrogen ortho phosphate buffer (40:60 v/v) at a flow rate of 1.0 ml/min. The method showed linear in the mentioned concentrations having line equation y = 22.848x + 1.3214 with correlation coefficient R2 of 0.9995. The recovery values for Ticagrelor ranged from 99.63% to 100.34%. The % RSD was 0.49% and 0.54%, respectively for intraday and interday precision. The limit of detection and limit of quantification were 0.05μg/mL and 0.20μg/mL respectively. Newly developed method was statistically validated for accuracy, precision, linearity and solution stability; hence it is directly applicable for the estimation of Ticagrelor up to trace level in routine analysis.
URI: https://ijpsr.com/bft-article/an-improved-assay-method-for-the-estimation-of-ticagrelor-hydrochloride-by-reverse-phase-liquid-chromatography/
http://10.9.150.37:8080/dspace//handle/atmiyauni/677
ISSN: 0975-8232
Appears in Collections:01. Journal Articles

Files in This Item:
File Description SizeFormat 
22-Vol.-7-Issue-5-May-2016-IJPSR-RA-6131.pdf375 kBAdobe PDFView/Open
Show full item record


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.